FORS1 - blue upgrade
A new CCD mosaic detector with blue optimised E2V detectors for FORS1 has been successfully commissioned the first week of April 2007. It is then offered for P79 (since April 7, 2007).Users preparing their Phase I package for the P80 Call for Proposals, should use the information given below in order to best take advantage of the greatly increased blue sensitivity of the system. Please note that this includes the new high through-put filters which will replace the old Bessel filters from the beginning of P80.
We have prepared a phase 2 package for operations with the new E2V mosaic detector. Users are requested to confirm the validity of the modified package considering the informations which are collected in the following fact sheet.
Users may want to review:
- Offset strategies considering the fringing.
- Exposure times considering the lower readout noise which may allow to take more exposures with shorter exposure times.
- Filter selection considering the new high throughput cutoff filters
- Target positions considering the gap between the two detectors about 33" South (for rotator angle 0 degree) of the nominal pointing position.
- Users which require high S/N ratios at >650nm should contact us (if they were not yet contacted)
- The use of un-binned readout mode, if the reduced sampling (0.25"/pxl in default 2x2 binned modes) appears problematic, would require a waiver for authorisation.
- Windowing will no longer be offered with the new CCD, as is currently the case in FORS2.
detector parameters
| default readout | imaging | 200KHz,2x2,low |
| spectroscopy | 100KHz,2x2,high | |
| read out noise | 200KHz,2x2,low (chip 1,2) | 4.16, 4.03 e- |
| 100KHz,2x2,high (chip 1,2) | 3.18, 2.78 e- | |
| gain factors | high gain (chip 1,2) | 0.54, 0.58 e-/adu |
| low gain (chip 1,2) | 2.22, 2.37 e-/adu | |
| image scale | binning 2x2, COLL_SR | 0.25"/pxl |
| binning 1x1, COLL_SR | 0.125"/pxl | |
| binning 2x2, COLL_HR | 0.125"/pxl |
Fringes at >650nm
The fringe amplitude of the new E2V detector at >650nm is larger then for the Tek 2K CCD.- The fringes will remain in the night
sky background in imaging modes with filters of a central wavelength > 650nm
even after the flat fielding of the data:
For the further data reduction it will be required to subtract the scaled night sky background from the frames.
- In spectroscopic modes only a partial correction of the fringe
can be obtained from the flat fielding:
- At wavelength >650nm, signal noise ratio of >15 may not be obtained due to the residual of the fringe corrections. Sky subtraction at these wavelengths will require that the target is observed at offset positions (nodding on the slit). This will however not correct for the residual fringes in the extracted science spectra.
Instrument response
There is a great improvement in quantum efficiency at UV (0.8 mag) and blue (0.4 mag) wavelength range as seen in the following table:
| filter | zero point (e-/s) | colour term | gain |
| U_BESS | 26.033+-0.045 | 0.091+-0.015 | +0.83 |
| B_BESS | 28.112+-0.034 | -0.107+-0.016 | +0.41 |
| V_BESS | 28.214+-0.022 | 0.020+-0.010 | +0.16 |
| R_BESS | 27.962+-0.022 | negligible | -0.04 |
| I_BESS | 26.993+-0.028 | -0.060+-0.024 | -0.16 |
The increased response of the new mosaic in the blue part of the spectrum can be seen also in the spectroscopic efficiencies measured for grisms #600B and #300V:
New filters
There are in addition new filter available which can either be used in the ubvRIz sequence or the ugrIz sequence to cover the full wavelength range with broad band filters:
Some filters like u_, b_, v_, g_HIGH and B_BESS are only available with FORS1 and others like U_, B_, V_BESS and R_SPECIAL only with FORS2. With the new FORS1 filters we have obtained the following photometric zero points:
| filter | zero point (e-/s) | colour term | gain |
| u_HIGH | 26.525+-0.051 | 0.106+-0.017 | +1.33 |
| b_HIGH | 28.480+-0.030 | 0.069+-0.014 | +0.78 |
| v_HIGH | 28.332+-0.022 | 0.049+-0.010 | +0.28 |
| g_HIGH | 28.888+-0.026 | 0.031+-0.012 |
