QC documentation system: QC procedure plot_data_dispersed.csh for MIDI

This documentation is intended both for QC scientists and SciOps astronomers (who may want to ignore the technical information displayed in grey).
 
HOME
 
CRIRES
FORS2
 
FORS1
GIRAFFE
UVES
XSHOOTER
 
ISAAC
VIMOS
VISIR
 
HAWKI
NACO
SINFONI
 
AMBER
MIDI
 
OMEGACAM
VIRCAM
[best viewing with style sheets and javascript enabled]
run_data_quality_processQC.sh
extractVis_dispersed.csh extractTF_dispersed.csh extractbinned_dispersed.csh
plot_data_dispersed.csh midi_frg_hs_disp_display.prg midi_frg_sp_disp_display.prg
 
bottom  plots   |   contact
NAME plot_data_dispersed.csh
VERSION 1.0 -- 12-06 implement docu
SYNTAX SHELL
CALL processQC -a abname or
. $DFO_PROC_DIR/run_data_quality_processQC.csh $RAW_TYPE $DATE $PF
. $DFO_PROC_DIR/plot_data_dispersed.csh $MODE $RT $DATE $PF $TRF
INSTRUMENT MIDI
RAWTYPE CAL_HS_FRT
CAL_SP_FRT
SCI_HS_FRT
SCI_SP_FRT
PURPOSE plot dispersed Vis and TF using the previously extracted values (extracted by extractxx_dispersed.csh)
PROCINPUT products under CAL/SCI_HS/SP_FRT
QC1TABLE trending | table(s) in QC1 database:
midi_calibrators | midi_TransferFunction
TRENDPLOT trending | HealthCheck plot associated to this procedure: none
QC1PAGE trending | associated documentation:
fullTF_qc1.html
QC1PLOTS
top

product_Vis.gif
Dispersed Visibility
product_TF.gif
Dispersed Transfer_Function
QC1PARAM
ALGORITHM The Visibility and TF are plotted for the full spectrum (170 channels for PRISM and 260 for GRISM)
CERTIF (guidelines for acceptance/rejection): the TF should be smooth accross the full spectrum. For most of the baseline the TF is around .6/.8
The Visibility should be smooth accross the spectrum for a calibrator and larger than .4. If is a science, larger than .075
COMMENTS (any comments)