# number of points on sci filter + aux + bad z = 2
# number of points on sci filter               = 2
# number of positions in lambda                = 301
# noof scans  1            1          
# x_pos     162.640000  -162.640000 
# y_pos       0.000000     0.000000 
# angle       2.960000     2.960000 
# z_pos      10.000000    10.000000 
# r_pos     162.640000   162.640000 
# phi         0.000000     3.141593 
# gnupl col      2            3      
# pos_id        49           50      
#
# lambda for peak transmission
#          567.0320000  567.0320000 
# peak transmission
#            0.7003849    0.6989855 
# lambda blue with 1/2 of peak transmission
#          502.7648834  502.2968043 
# lambda red  with 1/2 of peak transmission
#          600.6006265  599.5890518 
# half width                              
#           97.8357431   97.2922475 
# central wavelength                      
#          551.6827550  550.9429281 
# central wavelength calculated as SAGEM (mean frequency)
#          547.3451930  546.6476640 
# lambda blue with 80% of peak transmission
#          506.0510942  505.7732116 
# lambda red  with 80% of peak transmission
#          597.8106507  597.0612121 
# lambda blue with 0.0001 of peak transmission
#          478.0215856  477.7565204 
# lambda red  with 0.0001 of peak transmission
#          626.5113710  625.1601920 
# ATB calculated from points above 80%
#            0.6590747    0.6549962 
# ATB calculated from points above 50%
#            0.6449283    0.6505169 
# ABR calculated from points below and above 0.0001
#            0.0000431    0.0000321 
# minCWL      = 550.942928 at column =  3, at id = 50 
# maxCWL      = 551.682755 at column =  2, at id = 49 
# difference =    0.7398269 
# minBW       = 97.292248 at column =  3, at id = 50 
# maxBW       = 97.835743 at column =  2, at id = 49 
# difference =    0.5434956 
# minATB80    = 0.654996 at column =  3, at id = 50 
# maxATB80    = 0.659075 at column =  2, at id = 49 
# difference =    0.0040784 
# minBlueEdge = 502.296804 at column =  3, at id = 50 
# maxBlueEdge = 502.764883 at column =  2, at id = 49 
# difference =    0.4680791 
# minRedEdge  = 599.589052 at column =  3, at id = 50 
# maxRedEdge  = 600.600627 at column =  2, at id = 49 
# difference =    1.0115747 
# minABR      = 0.000032 at column =  3, at id = 50 
# maxABR      = 0.000043 at column =  2, at id = 49 
# difference =    0.0000109 
  299.645    -0.000191     0.000014 
  302.654    -0.000051    -0.000100 
  305.662    -0.000026     0.000002 
  308.671     0.000019     0.000017 
  311.679    -0.000015    -0.000003 
  314.687     0.000002    -0.000006 
  317.694     0.000013    -0.000008 
  320.702     0.000031     0.000001 
  323.709     0.000019    -0.000014 
  326.716     0.000011    -0.000010 
  329.665     0.000014    -0.000005 
  332.672     0.000016     0.000005 
  335.678     0.000004    -0.000001 
  338.683     0.000010    -0.000001 
  341.689     0.000008    -0.000004 
  344.694     0.000004     0.000002 
  347.699     0.000003     0.000003 
  350.703     0.000003    -0.000001 
  353.707     0.000007    -0.000001 
  356.711     0.000005    -0.000003 
  359.715     0.000003    -0.000001 
  362.718     0.000003    -0.000001 
  365.721     0.000004    -0.000003 
  368.724     0.000007     0.000002 
  371.726     0.000007    -0.000000 
  374.728     0.000004    -0.000001 
  377.730     0.000004    -0.000000 
  380.788     0.000004     0.000001 
  383.789     0.000004    -0.000001 
  386.789     0.000001     0.000000 
  389.790     0.000002     0.000001 
  392.790     0.000002     0.000000 
  395.790     0.000001     0.000000 
  398.789     0.000001     0.000001 
  401.788     0.000003    -0.000001 
  404.787     0.000002     0.000000 
  407.785     0.000002    -0.000001 
  410.783     0.000002     0.000001 
  413.781     0.000002     0.000000 
  416.834     0.000002    -0.000000 
  419.831     0.000001    -0.000000 
  422.828     0.000002     0.000000 
  425.824     0.000002     0.000001 
  428.820     0.000001     0.000000 
  431.815     0.000002     0.000000 
  434.810     0.000002     0.000001 
  437.862     0.000001    -0.000000 
  440.856     0.000001     0.000000 
  443.850     0.000001     0.000000 
  446.844     0.000001    -0.000000 
  449.837     0.000002     0.000000 
  452.830     0.000002    -0.000000 
  455.878     0.000002     0.000001 
  458.870     0.000001     0.000000 
  461.862     0.000002     0.000001 
  464.853     0.000002     0.000001 
  467.901     0.000002     0.000001 
  470.891     0.000017     0.000017 
  473.881     0.000034     0.000033 
  476.871     0.000044     0.000050 
  479.916     0.000113     0.000117 
  482.905     0.000327     0.000370 
  485.894     0.000538     0.000540 
  488.882     0.001342     0.001530 
  491.925     0.003363     0.004208 
  494.913     0.012677     0.015006 
  497.899     0.045107     0.057198 
  500.942     0.178172     0.215494 
  503.928     0.459952     0.510829 
  506.913     0.601049     0.589060 
  509.954     0.599951     0.598449 
  512.939     0.613975     0.610346 
  515.923     0.629460     0.627633 
  518.963     0.606267     0.603112 
  521.946     0.657167     0.666671 
  524.929     0.664824     0.663437 
  527.967     0.669424     0.672404 
  530.949     0.671520     0.669907 
  533.987     0.687150     0.685117 
  536.968     0.683342     0.683718 
  539.948     0.674915     0.669646 
  542.984     0.634986     0.634901 
  545.964     0.622338     0.617101 
  548.999     0.624024     0.622802 
  551.977     0.644654     0.637073 
  555.012     0.659565     0.657722 
  557.989     0.669598     0.660561 
  561.023     0.662988     0.659474 
  563.999     0.683373     0.688042 
  567.032     0.700385     0.698986 
  570.007     0.676931     0.670747 
  573.038     0.676806     0.680153 
  576.013     0.686063     0.680461 
  579.043     0.690272     0.691611 
  582.017     0.696903     0.694860 
  585.046     0.692151     0.690172 
  588.019     0.690429     0.686073 
  591.047     0.678504     0.671057 
  594.019     0.666440     0.664553 
  597.046     0.617119     0.560450 
  600.073     0.392223     0.309347 
  603.043     0.155635     0.100472 
  606.068     0.036560     0.024616 
  609.038     0.011904     0.007791 
  612.062     0.003437     0.002448 
  615.086     0.001388     0.000956 
  618.053     0.000475     0.000348 
  621.076     0.000241     0.000186 
  624.099     0.000115     0.000081 
  627.064     0.000060     0.000050 
  630.086     0.000042     0.000035 
  633.106     0.000024     0.000017 
  636.071     0.000011     0.000006 
  639.090     0.000010     0.000010 
  642.109     0.000010     0.000006 
  645.128     0.000001     0.000001 
  648.090     0.000002     0.000001 
  651.107     0.000001     0.000001 
  654.123     0.000002     0.000001 
  657.139     0.000001     0.000001 
  660.099     0.000001     0.000001 
  663.114     0.000001     0.000001 
  666.129     0.000001     0.000000 
  669.142     0.000001     0.000000 
  672.155     0.000001     0.000001 
  675.168     0.000001    -0.000000 
  678.124     0.000001    -0.000000 
  681.136     0.000001     0.000001 
  684.146     0.000001     0.000001 
  687.157     0.000001     0.000001 
  690.166     0.000001     0.000000 
  693.175     0.000001     0.000001 
  696.183     0.000001     0.000001 
  699.191     0.000001     0.000000 
  702.198     0.000001     0.000001 
  705.205     0.000001     0.000001 
  708.211     0.000001     0.000001 
  711.216     0.000001     0.000000 
  714.220     0.000001     0.000001 
  717.224     0.000003     0.000001 
  720.228     0.000002     0.000001 
  723.230     0.000007     0.000014 
  726.232     0.000014     0.000012 
  729.234     0.000001     0.000004 
  732.234     0.000017     0.000029 
  735.235     0.000062     0.000056 
  738.234     0.000055     0.000062 
  741.233     0.000056     0.000029 
  744.231     0.000002     0.000000 
  747.228     0.000002     0.000001 
  750.225     0.000002     0.000001 
  753.221     0.000015     0.000014 
  756.272     0.000074     0.000130 
  759.267     0.000262     0.000166 
  762.261     0.000022     0.000001 
  765.254     0.000002     0.000002 
  768.247     0.000001     0.000001 
  771.294     0.000001     0.000001 
  774.286     0.000001     0.000001 
  777.277     0.000016     0.000015 
  780.266     0.000020     0.000015 
  783.311     0.000014     0.000000 
  786.300     0.000002     0.000004 
  789.288     0.000001     0.000000 
  792.275     0.000002    -0.000000 
  795.316     0.000001    -0.000001 
  798.302     0.000000    -0.000000 
  801.287     0.000000    -0.000000 
  804.327     0.000000     0.000000 
  807.310     0.000000     0.000000 
  810.348     0.000001     0.000000 
  813.331     0.000001     0.000000 
  816.312     0.000001     0.000000 
  819.348     0.000000     0.000000 
  822.328     0.000001    -0.000000 
  825.362     0.000001    -0.000000 
  828.341     0.000001     0.000001 
  831.374     0.000000     0.000000 
  834.351     0.000001    -0.000000 
  837.382     0.000000     0.000000 
  840.357     0.000000     0.000000 
  843.387     0.000001    -0.000000 
  846.361     0.000001    -0.000000 
  849.389     0.000000    -0.000000 
  852.362     0.000001    -0.000000 
  855.389     0.000000    -0.000000 
  858.359     0.000001    -0.000000 
  861.385     0.000001    -0.000001 
  864.409     0.000001    -0.000001 
  867.378     0.000000    -0.000000 
  870.400     0.000000     0.000000 
  873.422     0.000000     0.000000 
  876.389     0.000000    -0.000000 
  879.409     0.000001    -0.000000 
  882.428     0.000000    -0.000000 
  885.447     0.000000    -0.000000 
  888.410     0.000001     0.000000 
  891.427     0.000001    -0.000000 
  894.444     0.000001    -0.000000 
  897.459     0.000001     0.000000 
  900.419     0.000000    -0.000000 
  903.433     0.000000    -0.000000 
  906.446     0.000000     0.000000 
  909.458     0.000000     0.000000 
  912.469     0.000000     0.000000 
  915.480     0.000000     0.000000 
  918.489     0.000000     0.000000 
  921.443     0.000000     0.000000 
  924.451     0.000000     0.000000 
  927.458     0.000000     0.000000 
  930.465     0.000000     0.000000 
  933.470     0.000000     0.000000 
  936.475     0.000000     0.000000 
  939.478     0.000000     0.000000 
  942.481     0.000000     0.000000 
  945.483     0.000000     0.000000 
  948.484     0.000000     0.000000 
  951.484     0.000000     0.000000 
  954.538    -0.000000     0.000000 
  957.537     0.000000     0.000000 
  960.534     0.000002     0.000001 
  963.531     0.000002     0.000002 
  966.526     0.000000     0.000000 
  969.521     0.000000     0.000000 
  972.515     0.000000    -0.000000 
  975.563    -0.000000    -0.000000 
  978.555    -0.000000    -0.000000 
  981.546    -0.000000    -0.000000 
  984.537    -0.000000    -0.000000 
  987.580     0.000000    -0.000000 
  990.569     0.000000    -0.000000 
  993.556     0.000000    -0.000000 
  996.543     0.000000     0.000000 
  999.583     0.000000     0.000000 
 1002.568     0.000000     0.000000 
 1005.552     0.000000     0.000000 
 1008.589     0.000000     0.000000 
 1011.571     0.000000     0.000000 
 1014.606     0.000000    -0.000000 
 1017.587    -0.000000    -0.000000 
 1020.620     0.000000    -0.000000 
 1023.598     0.000000    -0.000000 
 1026.629     0.000000    -0.000000 
 1029.606     0.000000    -0.000000 
 1032.635     0.000000     0.000000 
 1035.610     0.000000     0.000000 
 1038.637     0.000000     0.000000 
 1041.610    -0.000000    -0.000000 
 1044.635     0.000000     0.000000 
 1047.660    -0.000000     0.000000 
 1050.629    -0.000000     0.000000 
 1053.652     0.000000    -0.000000 
 1056.620     0.000000    -0.000000 
 1059.640    -0.000000    -0.000000 
 1062.660     0.000000    -0.000000 
 1065.678     0.000000    -0.000000 
 1068.642     0.000000    -0.000000 
 1071.658     0.000000    -0.000000 
 1074.674     0.000000    -0.000000 
 1077.688     0.000000    -0.000000 
 1080.702     0.000000    -0.000000 
 1083.660     0.000000     0.000000 
 1086.672     0.000000     0.000000 
 1089.682     0.000000     0.000000 
 1092.691     0.000000     0.000000 
 1095.699     0.000000     0.000000 
 1098.707     0.000000     0.000000 
 1101.713     0.000000     0.000000 
 1104.718     0.000000     0.000000 
 1107.722     0.000000     0.000000 
 1110.725     0.000000     0.000000 
 1113.727     0.000000     0.000000 
 1116.728     0.000000     0.000000 
 1119.728     0.000000     0.000000 
 1122.726     0.000000     0.000000 
 1125.724     0.000000     0.000000 
 1128.720     0.000000     0.000000 
 1131.770     0.000000     0.000000 
 1134.764     0.000000     0.000000 
 1137.757     0.000000     0.000000 
 1140.749     0.000000     0.000000 
 1143.741     0.000000     0.000000 
 1146.784     0.000000     0.000000 
 1149.773     0.000000     0.000000 
 1152.761     0.000000     0.000000 
 1155.801     0.000000     0.000000 
 1158.786     0.000000     0.000000 
 1161.771     0.000000     0.000000 
 1164.807     0.000000     0.000000 
 1167.790     0.000000     0.000000 
 1170.771     0.000000     0.000000 
 1173.804     0.000000     0.000000 
 1176.783     0.000000     0.000000 
 1179.814     0.000000     0.000000 
 1182.790     0.000000     0.000000 
 1185.819     0.000000     0.000000 
 1188.793     0.000000     0.000000 
 1191.820     0.000000     0.000000 
 1194.845     0.000000     0.000000 
 1197.815     0.000000     0.000000 
 1200.838     0.000000     0.000000 