Xenon-Argon List for ISAAC -------------------------- Notes ----- i) Blended at LR means that the line is blended with some other line at a resolution of 3000. This corresponds to the resolution of the LR grating with the 0.3" slit. Spectra taken with wider slits will have many more blened lines. ii) Blended at MR means that the lines is blended with some other line at a resolution of 10000. This corresponds to the resolution of the LR grating with the 0.3" slit. Spectra taken with wider slits will have many more blended lines. iii) Intensities were calculated from spectra taken with the LR grism at the standard settings. The standard LR settings are: 0.90, 1.06, 1.25, 1.65 and 2.20 microns. When lines appear in more than one setting, the intensity is reported for both settings. iv) Settings at 0.90, 1.06, 1.25 and 1.65 microns suffer from order overlap. The lines from different orders are reported in the last table. v) Some lines that lie outside the normal LR range are also reported. vi) The follwoing table lists the details of the LR dispersion solution. The residuals can be reduced by using only the brightest lines and avoiding blends. Order FWR EWR OSF Arc NOL Dispersion Residual --------------------------------------------------------------------- 6 7850-10300 8300-9700 Z Xenon 9 2.38 0.06 Argon 12 2.38 0.12 5 9230-12150 9800-11200 SZ Xenon 19 2.85 0.14 Argon 20 2.85 0.21 4 10780-14420 11000-13500 J Xenon 20 3.56 0.23 Argon 45 3.56 0.24 3 14210-19060 14200-18200 SH Xenon 15 4.74 0.25 Argon 30 4.74 0.19 2 18550-25850 19200-25500 SK Xenon 17 7.13 0.23 Argon 26 7.14 0.37 -------------------------------------------------------------------- OSF = Order sorting filter FWR = Full wavelength range EWR = Effective wavelength range.Includes filter, atmosphere and order overlap. NOL = number of lines used in the fit. vii) There are a few unidentified lines in the ISAAC spectra, but most are fainter than 1000 units. References ---------- 1 - Wavelengths and Transition Probabilities for Atoms and Atomic Ions, NSRDS-NBS 68, Reader, Corliss, Wiese and Martin. Converted from air to vacuum wavelengths. 2 - CASPIR User's Manual, P. McGregor (MSSSO). Converted from air to vacuum 3 - CRSP User's Manual (KPNO) - Already in vacuum wavelengths 4 High resolution Argon Atlas by Lloyd Wallace, Ken Hinkle and Dick Joyce (KPNO). Already in vacuum wavelengths See: http://www.noao.edu/kpno/phoenix/tharne/ Conversion from air to vaccuum wavelengths was done via the table in Astrophysical Quantities, Allen, 1973. Wavelength Line Source Strength Comments -------------------------------------------------------- 8266.79 ArI 1 1890 8282.39 XeI 1 1850 8349.11 XeI 1 540 8410.52 ArI 1 27660 8426.96 ArI 1 49500 8523.78 ArI 1 27100 8578.36 XeI 1 50 8608.14 ArI 1 50 8650.91 XeI 1 100 8670.32 ArI 1 4720 8694.59 XeI 1 -- Blend at LR 8699.25 XeI 1 -- Blend at LR 8741.79 XeI 1 320 8760.60 XeI 1 70 8821.83 XeI 1 90480 8864.75 XeI 1 560 8911.17 XeI 1 260 8933.28 XeI 1 360 8954.71 XeI 1 16350 8983.51 XeI 1 -- Blend at LR 8990.04 XeI 1 -- Blend at LR 9047.93 XeI 1 24120 9125.47 ArI 1 >140000 Saturated. Too bright to use. 9165.16 XeI 1 31220 9170.03 XeI 1 -- Blend at LR 9197.16 ArI 1 340 9227.03 ArI 1 33200 9294.08 ArI 1 320 9309.19 XeI 2 170 9356.78 ArI 1 2080 9377.33 XeI 1 780 9414.59 XeI 2 180 9447.93 XeI 2 280 9499.67 XeI 2 -- Blend at LR 9515.99 XeI 1 1390 9587.77 XeI 2 30 9660.43 ArI 1 35900 (Z) 11440 (SZ) Overlap b/n Z-SZ 9687.97 XeI 2 40 (Z) 160 (SZ) Overlap b/n Z-SZ 9703.65 XeI 2 20 9720.82 XeI 1 230 9787.18 ArI 1 12340 9802.38 XeI 1 120100 9925.91 XeI 1 >140000 Saturated. Too bright to use 10026.46 XeI 2 480 10054.81 ArI 1 300 10087.56 XeI 2 210 10110.13 XeI 2 1990 10128.23 XeI 2 150 10166.2 Ar 3 40 10191.16 XeI 2 90 10253.84 XeI 2 410 10311.96 ArI 2 50 10335.55 ArI 1 200 10472.92 ArI 1 47810 10480.90 ArI 1 -- Blended at LR 10487.68 XeI 2 80 10509.38 ArI 1 150 Reference 2 gives 10509.35 10530.74 XeI 2 1880 10676.49 ArI 1 11160 10684.70 ArI 1 -- Blended at LR 10703.9 Ar 3 130 10709.71 XeI 2 510 10736.81 ArI 1 70 10761.84 XeI 2 250 10762.10 ArI 1 80 10776.30 ArI 2 120 Overlap with J starts. 10841.34 Xe1 1 16390 (SZ) 1910 (J) Ref. 2 gives 10841.30 10883.94 ArI 2 490 (SZ) 130 (J) 10898.31 XeI 2 630 (SZ) 200 (J) 10953.74 ArI 2 550 (SZ) 250 (J) 11081.90 ArI 1 930 (SZ) 1090 (J) 11088.27 XeI 2 1120 (SZ) 1410 (J) 11109.50 ArI 2 110 (SZ) 150 (J) Uncertain ID 11130.2 XeI 2 230 (SZ) 360 (J) 11165.8 XeI 2 40 (SZ) 40 (J) 11312.7 XeI 2 25 (SZ) 40 (J) 11396.82 Ar 4 470 (SZ) 1450 (J) 11418.2 XeI 2 50 (SZ) 120 (J) 11444.97 ArI 4 180 (SZ) 11450 (J) 11470.69 ArI 4 40 (SZ) 690 (J) 11491.25 ArI 4 380 (SZ) 15000 (J) End of SZ overlap 11617.3 Xe 3 420 Beginning of Js 11671.90 ArI 4 8220 11722.70 ArI 4 2170 11736.44 ArI 4 -- Blended at LR 11745.5 Xe 3 3360 11796.8 Xe 3 700 11860.5 Xe 3 760 11887.74 ArI 4 -- Blended at LR 11899.89 ArI 4 -- Blended at LR 11915.5 XeI 2 140 11946.55 ArI 4 1580 Blended at MR, ArI 11947.14 11956.4 Xe 3 340 12029.94 ArI 4 1123 12088.1 Xe 3 460 12115.64 ArI 4 17600 12143.06 ArI 4 11310 12154.69 ArI 4 -- Blended at LR 12207.15 XeI 2 60 12238.59 XeI 1 1320 12261.11 XeI 1 220 12346.77 ArI 4 120 12359.67 ArI 4 9210 12406.22 ArI 4 12406 12422.79 ArI 4 -- Blended at LR 12442.73 ArI 4 57730 12459.52 ArI 4 12460 12491.08 ArI 4 12490 12557.76 ArI 4 600 12593.64 XeI 1 560 12599.66 ArI 4 120 12626.84 XeI 1 14150 12641.94 ArI 4 320 12705.75 ArI 4 17640 12736.90 ArI 4 5500 12749.72 ArI 4 4000 12806.24 ArI 4 18230 12936.73 ArI 4 8500 12960.20 ArI 4 67320 13011.82 ArI 4 15880 13031.99 ArI 4 850 13217.61 ArI 4 27410 13231.72 ArI 4 25290 13234.51 ArI 4 -- Blended at LR 13276.27 ArI 4 39640 13305.95 Ar 4 -- Blended at LR 13316.85 ArI 4 29210 13333.76 ArI 4 1310 13370.77 ArI 4 61870 13410.26 ArI 4 5230 13503.10 ArI 4 -- Blended at LR 13507.88 ArI 4 91140 13547.91 ArI 4 1450 13577.33 ArI 4 2740 13603.05 ArI 4 3480 13626.38 ArI 4 25460 13660.79 XeI 1 6460 13682.29 ArI 4 16020 End of Js 13722.3 Ar IRAF 107900 13829.5 Ar IRAF -- Blended at LR 13832.1 Ar IRAF -- Blended at LR 13914.4 Ar IRAF 630 14097.49 ArI 4 140 (J) Beginning of SH 14146.31 XeI 1 -- Beyond the normal LR range 14178.58 ArI 4 -- Beyond the normal LR range 14244.85 XeI 1 1790 14253.09 ArI 4 -- Blended at LR 14260.76 ArI 4 -- Blended at LR, and at MR 14258.04 ArI 14368.91 XeI 1 1070 14581.44 ArI 4 -- Blended at LR 14600.46 ArI 4 2210 Blended at MR with 14599.73 ArI 14638.42 ArI 4 -- Blended at LR 14654.35 ArI 4 -- Blended at LR 14664.81 XeI 1 360 14697.03 ArI 4 910 Blended at MR with 14697.69 ArI 14736.83 XeI 1 40000 14743.17 ArI 4 1700 14790.11 ArI 4 350 ID uncertain 14880.57 ArI 4 80 15034.62 ArI 4 -- Blended at LR, and at MR 15035.28 ArI 15050.62 ArI 4 28600 15103.84 XeI 1 280 15176.84 ArI 4 6320 15306.07 ArI 4 7720 15333.54 ArI 4 4800 15353.44 ArI 4 -- Blended at LR 15357.32 ArI 4 -- Blended at LR 15406.85 ArI 4 1570 15422.60 XeI 1 17110 15450.95 ArI 4 100 15495.20 XeI 2 110 15559.71 ArI 4 180 Blended at MR with XeI 15561.38 15561.38 XeI 1 420 Blended at MR with ArI 15559.71 15739.21 ArI 4 200 15797.47 ArI 4 100 15821.10 ArI 4 230 15887.51 ArI 4 -- Blended at LR 15904.03 ArI 4 2500 15983.90 XeI 1 1060 15993.86 ArI 4 8700 16044.28 XeI 1 -- Blended at LR 16057.66 XeI 1 6530 16127.06 ArI 4 1280 16184.44 ArI 4 1600 16268.51 ArI 4 250 16441.07 ArI 4 3140 16524.38 ArI 4 10400 16553.78 ArI 4 16553 Blended at MR with 16554.91 ArI 16559.01 XeI 1 360 16732.72 XeI 1 11520 16744.65 ArI 4 3450 16839.14 XeI 2 60 16887.68 XeI 2 90 16945.21 ArI 4 16945 17330.50 XeI 2 6350 17369.82 XeI 2 70 17406.67 ArI 4 190 17449.67 ArI 4 4900 17594.8 XeI 2 170 17828.87 ArI 1 890 17919.61 ArI 4 23690 End of SH 18353.01 ArI 4 -- Beyond the normal LR range 18423.09 ArI 4 -- Beyond the normal LR range 18432.80 ArI 4 -- Blended with 18434.81, 18433.43 ArI 18491.76 ArI 4 -- Beyond the normal LR range 18569.30 ArI 4 -- Blended at LR 18575.23 ArI 4 -- Blended at LR, + MR 18576.65 ArI 18637.37 ArI 4 330 18750.12 ArI 4 230 18793.26 XeI 1 2020 19129.04 ArI 4 30 19180.51 ArI 4 90 19822.91 ArI 4 3340 19950.52 ArI 4 -- Blended at LR Check ID and Wav. 19971.18 ArI 4 3900 20031.14 ArI 4 -- Blended at LR 20035.57 ArI 4 -- Blended at LR 20074.41 ArI 4 240 20192.70 XeI 1 840 20267.77 XeI 1 25930 20322.56 ArI 4 1990 20518.4 XeI 2 50 20574.43 ArI 4 450 20598.9 XeI 2 140 20621.86 ArI 4 31650 20652.77 ArI 4 890 20722.00 ArI 4 -- Blended at LR 20739.22 ArI 4 1200 Blended with 20740.10 20816.72 ArI 4 820 20890.3 XeI 2 50 20991.84 ArI 4 22440 21041.57 ArI 4 140 21116.7 XeI 2 80 21338.71 ArI 4 2500 21378.90 XeI 2 660 21475.94 XeI 1 4640 21540.09 ArI 4 10150 22045.58 ArI 4 3500 22083.21 ArI 4 12942 22118.66 ArI 4 360 22275.91 XeI 2 210 22412.93 XeI 2 810 22539.74 ArI 2 40 22624.45 XeI 2 390 22884.1 XeI 2 80 22971.04 XeI 2 160 22978.37 ArI 4 150 23028.70 XeI 2 100 23079.75 XeI 2 240 23111.56 XeI 2 50 23139.52 ArI 4 11910 23199.65 XeI 1 18000 23285.89 XeI 1 700 23450.3 XeI 2 160 23802.95 XeI 2 330 23851.54 ArI 4 10480 23973.06 ArI 4 13070 24450.4 XeI 3 1450 24709.05 XeI 2 490 24783.37 ArI 4 150 24831.48 XeI 1 44600 25132.13 ArI 4 5760 25152.69 XeI 1 2500 25166.3 XeI 3 -- Blended at LR 25512.18 ArI 4 3750 25668.02 ArI 4 900 26276.24 XeI 1 -- Beyond the normal LR range --------------------------------------------------------------------- Main lines of other orders. ============================================== Line Lamp OSF Order Location Intensity ----------------------------------------- 9447.93 Xe Z 5 7873 760 9515.99 Xe Z 5 7930 3560 9660.43 Ar Z 5 8050 42020 8283.39 Xe Z 7 9664 370 8349.11 Xe Z 7 9740 200 8411 Xe Z 7 9816 400 8410.52 Ar Z 7 9812 8700 8426.96 Ar Z 7 9831 14703 8523.78 Ar Z 7 9944 6490 8670.32 Ar Z 7 10115 860 9660.43 Ar SZ 6 11593 3420 9787.97 Ar SZ 6 11745 3410 9802.38 Xe SZ 6 11763 36925 9925.38 Xe SZ 6 11910 40624 11081.90 Ar J 5 13852 700 11396.82 Ar J 5 14246 660 11444.97 Ar J 5 14306 780 11470.69 Ar J 5 14338 280 11491.25 Ar J 5 14364 4810 11088.27 Xe J 5 13860 880 11130.2 Xe J 5 13913 200 14097.49 Ar SH 4 18797 6700 14253.09 Ar SH 4 19004 1860 Blended with 14253.09 Ar 14260.76 Ar SH 4 19014 1860 Blended with 14260.76 Ar ----------------------------------------------------------------