Components with slight outgassing

Small black internal connectors     (Harwin Data Mate)

Small black internal connectors

Because of their tiny size, we inserted six of these devices onto the probe holder. A slight emission, probably related to the black plastic, was detected at:

Line [amu] Fragment ions
30 C2H6+
31 CF+, CH2OH+
55 C4H7+
56 C4H8+ ?

At higher temperatures, the lines of 55 amu and 56 amu appeared also in the background. The utilization of the internal connectors for the heating of the probe holder may serve as an explanation for that.