[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
VIMOS trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2018-04-04T03:32:40 (UT)
  | now: 2024-12-22T17:22:41 (UT) muc03 QC pipeline: vimos-2.9.19 (installed 2016-03-17)
same group: LR_blue LR_red MR HR_blue HR_orange HR_red
close window history: 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017   2018   HEALTH
  ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 0.0876 pix VAL0.02,0.12 216 ids_rms this | last_yr | all Q2, rms of dispersion solution
2 QC1DB MEDIAN 0.0908 pix VAL0.02,0.11 208 ids_rms this | last_yr | all Q1, rms of dispersion solution
3 QC1DB MEDIAN 0.078 pix VAL0.02,0.11 216 ids_rms this | last_yr | all Q3, rms of dispersion solution
4 QC1DB MEDIAN 0.0823 pix VAL0.02,0.11 215 ids_rms this | last_yr | all Q4, rms of dispersion solution
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 2|
data source:vimos_mos_wave
(QC1 database)
dataset:ids_rms
median:0.0876pix
fixed thresholds:0.02...0.12pix
N_data plotted:216
[click plot for closeup]
Plot 2
scores: 2|
data source:vimos_mos_wave
(QC1 database)
dataset:ids_rms
median:0.0908pix
fixed thresholds:0.02...0.11pix
N_data plotted:208
[click plot for closeup]
Plot 3
scores: 2|
data source:vimos_mos_wave
(QC1 database)
dataset:ids_rms
median:0.078pix
fixed thresholds:0.02...0.11pix
N_data plotted:216
[click plot for closeup]
Plot 4
scores: 2|
data source:vimos_mos_wave
(QC1 database)
dataset:ids_rms
median:0.0823pix
fixed thresholds:0.02...0.11pix
N_data plotted:215
[click plot for closeup]
This plot

Trending plot for rms of dispersion solution from MOS arc lamp measurements.

Only arc lamp measurements taken to calibrate MOS science observations are used; exposures with the standard star masks are disregarded.

In case of masks with only one science slit, the pipeline may report an artificially low value for the rms. Such values do very likely not indicate an instrument issue, scoring thresholds have been adapted to this.

Template ID: mos_cal_Arc (daytime), mos_cal_NightCalib (attached night-time). Pipeline recipe: vmspcaldisp.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up