Plot ? |
Symb ? |
Source * |
Average ? |
Thresholds ? |
N_ data |
QC1 parameter |
Data downloads |
Remarks |
method |
value |
unit |
method |
value |
1 |
| QC1DB |
MEDIAN |
0.379 |
ADU/pix/hr |
VAL | 0.,0.7 |
15 |
qc_dark_current |
this |
last_yr |
all
|
DARK current averaged over all 32 detectors (ADU/pixel/hour) |
2 |
| QC1DB |
MEDIAN |
0.239 |
ADU/pix/hr |
none | |
450 |
qc_dark_current |
this |
last_yr |
all
|
DARK current for each individual detector (ADU/pixel/hour) |
3 |
| QC1DB |
MEDIAN |
106 |
part/cm2/hr |
VAL | 100.,120. |
15 |
qc_particle_rate |
this |
last_yr |
all
|
particle rate averaged over all 32 detectors (particles/cm2/hour) |
4 |
■ | LOCAL |
none |
|
none |
none | |
465 |
det_id |
n/a |
score per detector: GREEN = within the thresholds |
4 |
■ | LOCAL |
none |
|
none |
none | |
0 |
det_id |
n/a |
score per detector: RED = beyond the thresholds |
|
*Data sources: QC1DB: QC1 database; LOCAL: local data source
|
Plot 1 scores: not implemented
| data source: | omegacam_dark (QC1 database) |
dataset: | qc_dark_current | |
median: | 0.379 | ADU/pix/hr |
fixed thresholds: | 0.0...0.7 | ADU/pix/hr |
N_data plotted: | 15 |
[click plot for closeup] |
Plot 2 scores: not implemented
| data source: | omegacam_dark (QC1 database) |
dataset: | qc_dark_current | |
median: | 0.239 | ADU/pix/hr |
thresholds: | none | |
N_data plotted: | 450 |
[click plot for closeup] |
Plot 3 scores: not implemented
| data source: | omegacam_dark (QC1 database) |
dataset: | qc_particle_rate | |
median: | 106 | part/cm2/hr |
fixed thresholds: | 100.0...120.0 | part/cm2/hr |
N_data plotted: | 15 |
[click plot for closeup] |
Plot 4 scores: not implemented
| data source: | local data source
|
dataset: (numbers below apply to this dataset) | det_id | ■ |
average: | none |
thresholds: | none |
N_data plotted: | 465 |
[click plot for closeup] |
This plot
Trending plots for the detector dark current and particle rate.
QC parameters are derived from each of 32 detectors of OmegaCAM dark frames.
DB Name: omegacam_dark
Fits header: QC.DARK.CURRENT / QC.PARTICLE.RATE
QC1 DB: qc_dark_current / qc_particle_rate
Description:
The dark current and particle rate is measured from the pipeline product MEAN_DARK
Setup:
BIN.X / BIN.Y / READ.MODE / READ.SPEED = 1/1/normal/normal
These Health Check plots describe:
Fig. 1 dark_current_AVG: The dark current averaged over
all 32 detectors (in ADU/pixel/hour).
This plot is sensitive to any changes common to all detectors.
Fig. 2 gain_ALL: The dark current measured for each
individual detector (in ADU/pixel/hour). This plot is sensitive changes in individual
detectors.
Fig. 3 particle_rate_AVG: The dark frame particle rate
averaged over all 32 detectors (in particles/cm^2/hour).
Fig. 4 score_outliers: Scores of the dark current as measured for each of the 32 detectors.
Green = within the defined thresholds.
Red = beyond the defined thresholds.
OmegaCAM detector plane layout (detector_ID locations):
General information
Click on any of the plots to see a close-up version.
The latest date is indicated on top of the plot.
If configured,
- data points belonging to the latest date are specially marked
- statistical averages are indicated by a solid line, and thresholds by broken lines
- outliers are marked by a red asterisk. They are defined as data points outside the
threshold lines
- "aliens" (= data points outside the plot Y limits) are marked by a red arrow (↑ or ↓)
- you can download the data for each parameter set if the 'Data downloads' link shows up
|