[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
VIMOS trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2018-04-04T03:36:27 (UT)
  | now: 2024-12-23T00:01:05 (UT) muc03 QC pipeline: vimos-2.9.19 (installed 2016-03-17)
same group: LR_blue LR_red MR HR_blue HR_orange HR_red
close window history: 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017   2018   HEALTH
  ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEAN 6.57 pix VAL-10,10 35 trace2_slope this | last_yr | all Q2, IFU slope x 4096
2 QC1DB MEAN 0.652 pix VAL-10,10 35 trace2_slope this | last_yr | all Q1, IFU slope x 4096
3 QC1DB MEAN -6.19 pix VAL-10,10 35 trace2_slope this | last_yr | all Q3, IFU slope x 4096
4 QC1DB MEAN 1.36 pix VAL-10,10 35 trace2_slope this | last_yr | all Q4, IFU slope x 4096
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 6|
data source:vimos_ifu_score
(QC1 database)
dataset:trace2_slope
mean:6.57pix
fixed thresholds:-10...10pix
N_data plotted:35
[click plot for closeup]
Plot 2
scores: 6|
data source:vimos_ifu_score
(QC1 database)
dataset:trace2_slope
mean:0.652pix
fixed thresholds:-10...10pix
N_data plotted:35
[click plot for closeup]
Plot 3
scores: 6|
data source:vimos_ifu_score
(QC1 database)
dataset:trace2_slope
mean:-6.19pix
fixed thresholds:-10...10pix
N_data plotted:35
[click plot for closeup]
Plot 4
scores: 6|
data source:vimos_ifu_score
(QC1 database)
dataset:trace2_slope
mean:1.36pix
fixed thresholds:-10...10pix
N_data plotted:35
[click plot for closeup]
This plot

Trending plot for slope of fibre traces (IFU).

The QC1 parameter QC.IFU.TRACE2.SLOPE is multiplied with 4096 (the Y dimension of the CCD). The result should stay within +/- 10 pixels around zero.

Setting HR_orange is usually measured daily; other settings only when science frames have been observed.

Template ID: ifu_cal_DayCalib (daytime Health Checks), ifu_cal_NightCalib (attached night-time). Pipeline recipe: vmifucalib.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up