[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
VIMOS trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2018-04-04T03:36:09 (UT)
  | now: 2024-12-22T22:49:23 (UT) muc03 QC pipeline: vimos-2.9.19 (installed 2016-03-17)
same group: LR_blue LR_red MR HR_blue HR_orange HR_red
close window history: 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017   2018   HEALTH
  ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 0.168 pix VAL0,0.3 12 ids_rms this | last_yr | all Q2, rms of dispersion solution
2 QC1DB MEDIAN 0.163 pix VAL0,0.3 12 ids_rms this | last_yr | all Q1, rms of dispersion solution
3 QC1DB MEDIAN 0.167 pix VAL0,0.3 12 ids_rms this | last_yr | all Q3, rms of dispersion solution
4 QC1DB MEDIAN 0.169 pix VAL0,0.3 12 ids_rms this | last_yr | all Q4, rms of dispersion solution
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 1|
data source:vimos_ifu_score
(QC1 database)
dataset:ids_rms
median:0.168pix
fixed thresholds:0...0.3pix
N_data plotted:12
[click plot for closeup]
Plot 2
scores: 1|
data source:vimos_ifu_score
(QC1 database)
dataset:ids_rms
median:0.163pix
fixed thresholds:0...0.3pix
N_data plotted:12
[click plot for closeup]
Plot 3
scores: 1|
data source:vimos_ifu_score
(QC1 database)
dataset:ids_rms
median:0.167pix
fixed thresholds:0...0.3pix
N_data plotted:12
[click plot for closeup]
Plot 4
scores: 1|
data source:vimos_ifu_score
(QC1 database)
dataset:ids_rms
median:0.169pix
fixed thresholds:0...0.3pix
N_data plotted:12
[click plot for closeup]
This plot

Trending plot for the rms of the dispersion solution for IFU calibrations.

The values should show a stable trend without sudden outliers. Since the dispersion solution is usually very stable, any values above 0.3 to 0.5 pixels (depending on the setting) are a concern.

Since October 2010, values for the dispersion rms vary for Q3 and Q4 because of variable flux from the He and Ne lamps.

There are occasional pipeline failures in Q2 for daily Health Check calibrations of IFU LR_red; these give an artificially high value for the RMS of the dispersion solution.

Settings HR_orange and LR_red are usually measured daily; other settings only when science frames have been observed.

Template ID: ifu_cal_DayCalib (daytime Health Checks), ifu_cal_NightCalib (attached night-time). Pipeline recipe: vmifucalib.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up