UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-22T09:22:54 (UT)
  | now: 2024-11-22T10:51:48 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Median Master_RON Raw_RON Struct_X Struct_Y Pattern_noise
close window history: 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023   2024   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 0.0349 ADU VAL0.01,0.13 88 struct_col this | last_yr | all structure of master BIAS frame in Y direction, BLUE CCD, readout mode 225kHz,1x1,1pt,low
2 QC1DB MEDIAN 0.0915 ADU VAL0.03,0.13 106 struct_col this | last_yr | all same, for REDL CCD
3 QC1DB MEDIAN 0.209 ADU VAL0.16,0.26 106 struct_col this | last_yr | all same, for REDU CCD

4 QC1DB MEDIAN 0.238 ADU VAL0.20,0.30 86 struct_col this | last_yr | all same as Fig. 1, for readout mode 50kHz,2x2,high
5 QC1DB MEDIAN 0.128 ADU OFF± 0.05 96 struct_col this | last_yr | all same, for REDL CCD
6 QC1DB MEDIAN 0.279 ADU VAL0.22,0.34 96 struct_col this | last_yr | all same, for REDU CCD
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 9|
data source:uves_bias
(QC1 database)
dataset:struct_col
median:0.0349ADU
fixed thresholds:0.01...0.13ADU
N_data plotted:88
[click plot for closeup]
Plot 2
scores: 8|
data source:uves_bias
(QC1 database)
dataset:struct_col
median:0.0915ADU
fixed thresholds:0.03...0.13ADU
N_data plotted:106
[click plot for closeup]
Plot 3
scores: 8|
data source:uves_bias
(QC1 database)
dataset:struct_col
median:0.209ADU
fixed thresholds:0.16...0.26ADU
N_data plotted:106
[click plot for closeup]
Plot 4
scores: 8|
data source:uves_bias
(QC1 database)
dataset:struct_col
median:0.238ADU
fixed thresholds:0.2...0.3ADU
N_data plotted:86
[click plot for closeup]
Plot 5
scores: 9|
data source:uves_bias
(QC1 database)
dataset:struct_col
median:0.128ADU
rel. thresholds: 0.07825...0.17825ADU
N_data plotted:96
[click plot for closeup]
Plot 6
scores: 9|
data source:uves_bias
(QC1 database)
dataset:struct_col
median:0.279ADU
fixed thresholds:0.22...0.34ADU
N_data plotted:96
[click plot for closeup]
This plot

This is the trending report for the QC parameters derived from UVES BIAS frames.

These plots describe:

  • the collapsed structure in Y direction of the BLUE, RED LOWER and RED UPPER CCD master frame in the 1x1 binning, 225kHz, low gain readout mode;
  • the same for the 2x2 binning, 50kHz, high gain readout mode

Outliers in the HC plot "Struct_Y", #6 ("REDU 2x2 high") are due to remnants left by previous arc observations. Arcs taken in 760-860nm settings are known to leave this kind of remnances.

Find more information about UVES BIAS frames here.
Other QC parameters related to detector properties (like linearity, gain) are trended here.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up