[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SPHERE trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-23T04:30:09 (UT)
  | now: 2024-11-23T09:22:18 (UT) muc03 QC pipeline: spher-0.56.0 (installed 2024-09-26)
same group: med_master ron NumHotPixel
close window history: 2015 2016 2017 2018 2019 2020 2021 2022 2023   2024   HEALTH FULL
  ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN -0.465625 ADU VAL-1,2 21 qc_med this | last_yr | all DIT=2, NDIT=40, Nondest
2 QC1DB MEDIAN -0.554455 ADU VAL-1,2 21 qc_med this | last_yr | all DIT=8, NDIT=25, Nondest
3 QC1DB MEDIAN 1.04328 ADU VAL-1,4 21 qc_med this | last_yr | all DIT=30, NDIT=15, Nondest
4 QC1DB MEDIAN 78 degree none  576 dettemp this | last_yr | all IRDIS detector temperature, INS1.TEMP101.VAL
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 3|
data source:sphere_irdis_dark
(QC1 database)
dataset:qc_med
median:-0.465625ADU
fixed thresholds:-1...2ADU
N_data plotted:21
[click plot for closeup]
Plot 2
scores: 3|
data source:sphere_irdis_dark
(QC1 database)
dataset:qc_med
median:-0.554455ADU
fixed thresholds:-1...2ADU
N_data plotted:21
[click plot for closeup]
Plot 3
scores: 3|
data source:sphere_irdis_dark
(QC1 database)
dataset:qc_med
median:1.04328ADU
fixed thresholds:-1...4ADU
N_data plotted:21
[click plot for closeup]
Plot 4
scores: not implemented
data source:sphere_irdis_dark
(QC1 database)
dataset:dettemp
median:78degree
thresholds:none 
N_data plotted:576
[click plot for closeup]
This plot

This plot shows the IRDIS detector dark level values of the three regularly acquired health check setups

  • DIT=2, NDIT=40
  • DIT=8, NDIT=25
  • DIT=30, NDIT=15

Darks are used as a technical monitoring of the detector but are not used to calibrate science data.
The background frames are used to calibrate science frames from four IRDIS modes CI, DBI, DPI and LSS.

 


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up