[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SINFONI trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2019-06-28T05:02:02 (UT)
  | now: 2024-12-23T20:29:59 (UT) muc04 QC pipeline: sinfo-3.2.5 (installed 2019-05-09)
same group: lamp_eff   lamp_badpix lamp_off lamp_fpn1 lamp_fpn2 vignetting_check_sl1 vignetting_check_sl32
close window history: 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018   2019   HEALTH FULL
  ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 0.885 ADU VAL0.6,1.0 34 qc_rf32 this | last_yr | all relative flux (slitlet 32): J-band/250 mas scale
2 QC1DB MEDIAN 0.922 ADU VAL0.6,1.0 33 qc_rf32 this | last_yr | all relative flux (slitlet 32): H-band/250 mas scale
3 QC1DB MEDIAN 0.86 ADU VAL0.6,1.0 46 qc_rf32 this | last_yr | all relative flux (slitlet 32): K-band at the 250 mas camera scale
4 QC1DB MEDIAN 0.869 ADU VAL0.6,1.0 30 qc_rf32 this | last_yr | all relative flux (slitlet 32): H+K-band at the 250 mas camera scale

5 QC1DB MEDIAN 0.876 ADU VAL0.6,1.0 18 qc_rf32 this | last_yr | all relative flux (slitlet 32): J-band/ 25 mas scale
6 QC1DB MEDIAN 0.907 ADU VAL0.6,1.0 18 qc_rf32 this | last_yr | all relative flux (slitlet 32): H-band/ 25 mas scale
7 QC1DB MEDIAN 0.832 ADU VAL0.6,1.0 31 qc_rf32 this | last_yr | all relative flux (slitlet 32): K-band at the 25 mas camera scale
8 QC1DB MEDIAN 0.823 ADU VAL0.6,1.0 27 qc_rf32 this | last_yr | all relative flux (slitlet 32): H+K-band at the 25 mas camera scale

9 QC1DB MEDIAN 0.888 ADU VAL0.6,1.0 17 qc_rf32 this | last_yr | all relative flux (slitlet 32): J-band/100 mas scale
10 QC1DB MEDIAN 0.913 ADU VAL0.6,1.0 22 qc_rf32 this | last_yr | all relative flux (slitlet 32): H-band/100 mas scale
11 QC1DB MEDIAN 0.855 ADU VAL0.6,1.0 28 qc_rf32 this | last_yr | all relative flux (slitlet 32): K-band at the 100 mas camera scale
12 QC1DB MEDIAN 0.857 ADU VAL0.6,1.0 20 qc_rf32 this | last_yr | all relative flux (slitlet 32): H+K-band at the 100 mas camera scale

*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 2|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.885ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:34
[click plot for closeup]
Plot 2
scores: 5|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.922ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:33
[click plot for closeup]
Plot 3
scores: 5|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.86ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:46
[click plot for closeup]
Plot 4
scores: no current data
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.869ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:30
[click plot for closeup]
Plot 5
scores: 1|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.876ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:18
[click plot for closeup]
Plot 6
scores: 1|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.907ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:18
[click plot for closeup]
Plot 7
scores: 1|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.832ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:31
[click plot for closeup]
Plot 8
scores: no current data
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.823ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:27
[click plot for closeup]
Plot 9
scores: 1|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.888ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:17
[click plot for closeup]
Plot 10
scores: 1|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.913ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:22
[click plot for closeup]
Plot 11
scores: 4|
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.855ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:28
[click plot for closeup]
Plot 12
scores: no current data
data source:sinfoni_flat
(QC1 database)
dataset:qc_rf32
median:0.857ADU
fixed thresholds:0.6...1.0ADU
N_data plotted:20
[click plot for closeup]
This plot

These are the trending plots for the vignetting check computed from the LAMP FLATs.

Slitlets #1 and #32 correspond to the upper-most and the lower-most slitlets in the field-of-view.
As such, these slitlets can be subject to vignetting to the extent that their overall count rates can be a few 10% lower than for other slitlets. Following image reconstruction, these slitlets are located on the central part of the detector. A possible vignetting is monitored by comparing the relative count rates (i.e. the mean counts in the slitlet #1 or #32 divided by the mean counts in all other slitlets).
This QC1 parameter is most relevant following an instrument intervention.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up