PIONIER: Kappa Matrix
|
HC PLOTS |
interferometric transmission |
|
beam transmission |
|
QC1 database (advanced users):
browse |
plot
|
The internal flux ratio (interferometric transmission) is measured for the pixels on the detector (science outputs).
Data can be taken on sky (used to calibrate sky/interferometric data) or on the internal lamp (to monitor the instrument).
Four files are recorded, each of them has the shutter of one telescope opened (T), the other 3 beeing closed (F) : TFFF,FTFF, FFTF, FFFT.
The recipe computes the photometric transmission matrix. The first 40 elements of the scans are discarded and all values of the rest of the scans are averaged.
The normalization is done by averaging the flux of each telescope and per spectral channel.
KAPPA MATRIX
The internal flux ratio (interferometric transmission) is measured for the pixels on the detector (science outputs).
Data can be taken on sky (used to calibrate sky/interferometric data) or on the internal lamp (to monitor the instrument).
Four files are recorded, each of them has the shutter of one telescope opened (T), the other 3 beeing closed (F) : TFFF,FTFF, FFTF, FFFT.
QC1_parameters
FITS key |
QC1 database: table, name |
definition |
class* |
HC_plot** |
more docu |
QC.KAPPA12.AVG | pionier_kappa_matrix..kappa_avg12 | interferometric transmission bas 12 | HC | | [docuSys coming] |
QC.KAPPA13.AVG | pionier_kappa_matrix..kappa_avg13 | interferometric transmission bas 13 | HC | | [docuSys coming] |
QC.KAPPA14.AVG | pionier_kappa_matrix..kappa_avg14 | interferometric transmission bas 14 | HC | | [docuSys coming] |
QC.KAPPA23.AVG | pionier_kappa_matrix..kappa_avg23 | interferometric transmission bas 23 | HC | | [docuSys coming] |
QC.KAPPA24.AVG | pionier_kappa_matrix..kappa_avg24 | interferometric transmission bas 24 | HC | | [docuSys coming] |
QC.KAPPA34.AVG | pionier_kappa_matrix..kappa_avg34 | interferometric transmission bas 34 | HC | | [docuSys coming] |
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one. |
Trending
The value for the kappa matrix is plotted for the internal lamp and for sky data .
the 2 different plots can be found in the internal kappa
plot and in the "sky data" plot.
The Dark is plotted for the different set-up
Scoring&thresholds KAPPA MATRIX
The value of the thresholds have been defined together with the Instrument scientist
History
The evolution of the kappa matrix is shown here
Algorithm KAPPA MATRIX
The recipe computes the photometric transmission matrix. The first 40 elements of the scans are discarded and all values of the rest of the scans are averaged.
The normalization is done by averaging the flux of each telescope and per spectral channel.
~
|