[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
OMEGACAM trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-06-25T16:00:12 (UT)
  | now: 2024-06-26T05:26:18 (UT) muc06 QC pipeline: omega-1.1.8 (installed 2023-06-22)
same group: BIAS_level BIAS_readnoise RON_select BIAS_select BIAS_hotpixels
close window history: 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023   2024   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
    These plots monitor the mean level of the two raw bias frames that make up the read-noise template. The intent is to monitor OmegaCAM's troublesome detector ESO_CCD_#82 (red points) whose offset can fluctuate strongly, particularly following periods when the instrument has been idle.
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 243 ADU VAL230.,300. 118 qc_raw_bias1_mean this | last_yr | all mean level of first raw bias for detector ESO_CCD_#82 (ADU). HC scores only apply to this detector.
1 QC1DB none   ADU none  118 qc_raw_bias1_mean this | last_yr | all mean level of first raw bias for the stable detector ESO_CCD_#81 (ADU)
1 QC1DB none   ADU none  118 qc_raw_bias1_mean this | last_yr | all mean level of first raw bias for the stable detector ESO_CCD_#83 (ADU)
2 QC1DB MEDIAN 242 ADU VAL230.,300. 118 qc_raw_bias2_mean this | last_yr | all mean level of second raw bias for detector ESO_CCD_#82 (ADU). HC scores only apply to this detector.
2 QC1DB none   ADU none  118 qc_raw_bias2_mean this | last_yr | all mean level of second raw bias for the stable detector ESO_CCD_#81 (ADU)
2 QC1DB none   ADU none  118 qc_raw_bias2_mean this | last_yr | all mean level of second raw bias for the stable detector ESO_CCD_#83 (ADU)
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 15|
data source:omegacam_readnoise
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_raw_bias1_mean
median:243ADU
fixed thresholds:230.0...300.0ADU
N_data plotted:118
[click plot for closeup]
Plot 2
scores: 15|
data source:omegacam_readnoise
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_raw_bias2_mean
median:242ADU
fixed thresholds:230.0...300.0ADU
N_data plotted:118
[click plot for closeup]
This plot

Trending plots for the bias read-out noise. The intent is to monitor OmegaCAM's troublesome detector ESO_CCD_#82 whose offset can fluctuate strongly, particularly following periods when the instrument has been idle.

For comparison, the stable detectors ESO_CCD_#81 and ESO_CCD_#83 are plotted for reference.

DB Name: omegacam_readnoise
Fits header: QC.BIAS1.RAWREADNOISE
QC1 DB: qc_raw_bias1_mean / qc_raw_bias2_mean

Description:
The level of the two raw bias frames is measured

Setup:
BIN.X / BIN.Y / READ.MODE / READ.SPEED = 1/1/normal/normal

These Health Check plots describe:
Fig. 1 raw_bias1_mean: The mean level of the first raw bias frame is plotted for detectors #82, #81, and #83 (in ADU).
Fig. 2 raw_bias2_mean: The mean level of the second raw bias frame is plotted for detectors #82, #81, and #83 (in ADU).


OmegaCAM detector plane layout (detector_ID locations):
OmegaCAM detector layout

 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up