[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
NACO trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2019-10-07T09:27:38 (UT)
  | now: 2024-11-22T22:29:40 (UT) muc04 QC pipeline: naco-4.4.8 (installed 2019-03-26)
same group: RON_DAILY dark(DIT)   UNCORR
close window history: 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018   2019   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 5.13 ADU VAL3,10 83 dark_ron12 this | last_yr | all Double ReadResetRead DIT=0.5
2 QC1DB MEDIAN 11.1 ADU VAL5,20 83 dark_ron12 this | last_yr | all FowlerNSample DIT=300s
3 QC1DB MEDIAN 1.92 ADU VAL1.5,2.5 84 dark_ron12 this | last_yr | all FowlerNSample DIT=1.7892s
4 QC1DB MEDIAN 26.7 ADU VAL20,40 83 dark_median this | last_yr | all Double ReadResetRead DIT=0.5
5 QC1DB MEDIAN 868 ADU VAL300,1400 83 dark_median this | last_yr | all FowlerNSample DIT=300s
6 QC1DB MEDIAN 8.59 ADU VAL3,16 84 dark_median this | last_yr | all FowlerNSample DIT=1.7892s
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 8|
data source:naco_dark
(QC1 database)
dataset:dark_ron12
median:5.13ADU
fixed thresholds:3...10ADU
N_data plotted:83
[click plot for closeup]
Plot 2
scores: 8|
data source:naco_dark
(QC1 database)
dataset:dark_ron12
median:11.1ADU
fixed thresholds:5...20ADU
N_data plotted:83
[click plot for closeup]
Plot 3
scores: 8|
data source:naco_dark
(QC1 database)
dataset:dark_ron12
median:1.92ADU
fixed thresholds:1.5...2.5ADU
N_data plotted:84
[click plot for closeup]
Plot 4
scores: 8|
data source:naco_dark
(QC1 database)
dataset:dark_median
median:26.7ADU
fixed thresholds:20...40ADU
N_data plotted:83
[click plot for closeup]
Plot 5
scores: 8|
data source:naco_dark
(QC1 database)
dataset:dark_median
median:868ADU
fixed thresholds:300...1400ADU
N_data plotted:83
[click plot for closeup]
Plot 6
scores: 8|
data source:naco_dark
(QC1 database)
dataset:dark_median
median:8.59ADU
fixed thresholds:3...16ADU
N_data plotted:84
[click plot for closeup]
This plot

This plot shows the dark level and the readout noise for three acquired dark calibration setups, triggered by the daily HC batch (DET.NCORRS.NAME, DET.MODE.NAME, DET.DIT in sec). Values with DET.WIN.NX = 1024 and DET.FRAM.TYPE = INT are monitored only, meaning window read out mode and cube mode are excluded. No selection on the camera (e.g. S13, S27, ...) is applied.

  • Double_ReadResetRead, HighSensitivity, 0.5
  • FowlerNsamp, HighSensitivity, 300.0
  • FowlerNSamp, HighSensitivity, 1.7892
 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up