Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN -1.87 NONE none  60 A2 this | last_yr | all A2_OPEN (not scored)
2 QC1DB MEDIAN -0.14 NONE none  59 A2 this | last_yr | all A2_GRISM (not scored)
3 QC1DB MEDIAN -16 NONE none  58 A2 this | last_yr | all A2_PRISM (not scored)
4 QC1DB MEDIAN 2354 NONE VAL2000,3100 60 A1 this | last_yr | all A1_OPEN
5 QC1DB MEDIAN 2344 NONE VAL2000,3000 59 A1 this | last_yr | all A1_GRISM
6 QC1DB MEDIAN 5797 NONE VAL5000,7200 58 A1 this | last_yr | all A1_PRISM
7 QC1DB MEDIAN 5400 NONE VAL4000,6000 60 A0 this | last_yr | all A0_OPEN
8 QC1DB MEDIAN 5241 NONE VAL4000,6000 59 A0 this | last_yr | all A0_GRISM
9 QC1DB MEDIAN 5028 NONE VAL4000,6000 58 A0 this | last_yr | all A0_PRISM
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:midi_detlin
(QC1 database)
dataset:A2
median:-1.87NONE
thresholds:none 
N_data plotted:60
[click on plot for closeup]
Plot 2
data source:midi_detlin
(QC1 database)
dataset:A2
median:-0.14NONE
thresholds:none 
N_data plotted:59
[click on plot for closeup]
Plot 3
data source:midi_detlin
(QC1 database)
dataset:A2
median:-16NONE
thresholds:none 
N_data plotted:58
[click on plot for closeup]
Plot 4
data source:midi_detlin
(QC1 database)
dataset:A1
median:2354NONE
fixed thresholds:2000...3100NONE
N_data plotted:60
[click on plot for closeup]
Plot 5
data source:midi_detlin
(QC1 database)
dataset:A1
median:2344NONE
fixed thresholds:2000...3000NONE
N_data plotted:59
[click on plot for closeup]
Plot 6
data source:midi_detlin
(QC1 database)
dataset:A1
median:5797NONE
fixed thresholds:5000...7200NONE
N_data plotted:58
[click on plot for closeup]
Plot 7
data source:midi_detlin
(QC1 database)
dataset:A0
median:5400NONE
fixed thresholds:4000...6000NONE
N_data plotted:60
[click on plot for closeup]
Plot 8
data source:midi_detlin
(QC1 database)
dataset:A0
median:5241NONE
fixed thresholds:4000...6000NONE
N_data plotted:59
[click on plot for closeup]
Plot 9
data source:midi_detlin
(QC1 database)
dataset:A0
median:5028NONE
fixed thresholds:4000...6000NONE
N_data plotted:58
[click on plot for closeup]
This plot

The linearity of the MIDI detector is an indicator if the quality of its response curve. The pipeline calculate the coefficients of the polynomial fit of the average pixel value (over a windowed area) of the detector for 3 different configurations (image, prism and grism). The polynomial fit is: I(t) = A_0 + A_1 * t + A_2 * t^2, where I is the detector level in ADU and t the time in millisecond.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up