[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
ISAAC trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2013-11-30T20:22:57 (UT)
same group: SW_Offset SW_Disp SW_RESPOW SW_ArcLamps SW_XCORR LW_Offset LW_Disp LW_ResPow   DISP(LAM) RESP(LAM)
close window history: 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013   HEALTH
  ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
  • 2007-04-25: migrated to trendPlotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
2 QC1DB none   ADU none  1 qc_pixoff this | last_yr | all grating offset in pixel for LW, LR, SH
3 QC1DB none   ADU none  26 qc_pixoff this | last_yr | all grating offset in pixel for LW, LR, SL
4 QC1DB none   ADU none  69 qc_pixoff this | last_yr | all grating offset in pixel for LW, MR, J+Block
5 QC1DB none   ADU none  23 qc_pixoff this | last_yr | all grating offset in pixel for LW, MR, SH
6 QC1DB none   ADU none  24 qc_pixoff this | last_yr | all grating offset in pixel for LW, MR, SL
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 2
data source:isaac_arc
(QC1 database)
dataset:qc_pixoff
average:none
thresholds:none
N_data plotted:1
[click on plot for closeup]
Plot 3
data source:isaac_arc
(QC1 database)
dataset:qc_pixoff
average:none
thresholds:none
N_data plotted:26
[click on plot for closeup]
Plot 4
data source:isaac_arc
(QC1 database)
dataset:qc_pixoff
average:none
thresholds:none
N_data plotted:69
[click on plot for closeup]
Plot 5
data source:isaac_arc
(QC1 database)
dataset:qc_pixoff
average:none
thresholds:none
N_data plotted:23
[click on plot for closeup]
Plot 6
data source:isaac_arc
(QC1 database)
dataset:qc_pixoff
average:none
thresholds:none
N_data plotted:24
[click on plot for closeup]
This plot

This is the trending plot for the ISAAC arc lamp calibrations of the LW-arm

The QC parameter is the offset between the nominal central wavelength (read from the fits header INS.GRAT.WLEN) and the central wavelength derived from the wavelength solution of the pipeline recipe (HIERARCH ESO QC WLEN) . The offset is given in pixel units. For all setups in the LW-arm: Both Argon and Xenon lamps are on.

These plots describe:
2. Offset for LW-arm, LR-grating, SH-filter.
3. Offset for LW-arm, LR-grating, SL-filter.

4. Offset for LW-arm, MR-grating, J+Block-filter.
5. Offset for LW-arm, MR-grating, SH-filter.
6. Offset for LW-arm, MR-grating, SL-filter..

Find more information about ISAAC QC parameters derived from arc frames here.

 


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up