[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+R to enforce refresh of scores and dates]
ISAAC trending system: HISTORY plot
Last update: 2010-12-30T20:21:30 (UT)
same group: SW_IMG SW_SPEC   LW_DCHB LW_DCLB LW_UCHB det_temp
close window history: 2001 2002 2003 2004 2005 2006 2007 2008 2009   2010   2011   2012 2013   HEALTH
  ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEAN 6.31637   VAL5,7.5 86 ron12 this | last_yr | all RON12 for DIT=0.28sec and DoubleCorrHighBias
2 QC1DB MEAN 6.30385   VAL5.7,7 86 ron23 this | last_yr | all RON23 for DIT=0.28sec and DoubleCorrHighBias
3 QC1DB MEAN 49.25681 ADU 3SIG± 0.62871 86 qc_darkmed this | last_yr | all dark current for DIT=0.28sec and DoubleCorrHighBias
4 QC1DB MEAN 0.03021   VAL0,0.08 86 qc_dark_stdev this | last_yr | all raw framedark RMS for DIT=0.28sec and DoubleCorrHighBias
5 QC1DB MEAN 6.19795   VAL5.3,7.3 88 ron12 this | last_yr | all RON12 for DIT=0.56sec and DoubleCorrHighBias
6 QC1DB MEAN 6.24223   VAL5.6,6.8 88 ron23 this | last_yr | all RON23 for DIT=0.56sec and DoubleCorrHighBias
7 QC1DB MEAN 52.05605 ADU 3SIG± 0.62736 88 qc_darkmed this | last_yr | all dark current for DIT=0.56sec and DoubleCorrHighBias
8 QC1DB MEAN 0.03105   VAL0,0.06 88 qc_dark_stdev this | last_yr | all raw framedark RMS for DIT=0.56sec and DoubleCorrHighBias
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:isaac_dark
(QC1 database)
parameter | dataset:ron12
mean:6.31637 
fixed thresholds:5.0...7.5 
N_data:86
[click on plot for closeup]
Plot 2
data source:isaac_dark
(QC1 database)
parameter | dataset:ron23
mean:6.30385 
fixed thresholds:5.7...7.0 
N_data:86
[click on plot for closeup]
Plot 3
data source:isaac_dark
(QC1 database)
parameter | dataset:qc_darkmed
mean:49.25681ADU
rel. thresholds: 3σ48.6281...49.88552ADU
N_data:86
[click on plot for closeup]
Plot 4
data source:isaac_dark
(QC1 database)
parameter | dataset:qc_dark_stdev
mean:0.03021 
fixed thresholds:0.0...0.08 
N_data:86
[click on plot for closeup]
Plot 5
data source:isaac_dark
(QC1 database)
parameter | dataset:ron12
mean:6.19795 
fixed thresholds:5.3...7.3 
N_data:88
[click on plot for closeup]
Plot 6
data source:isaac_dark
(QC1 database)
parameter | dataset:ron23
mean:6.24223 
fixed thresholds:5.6...6.8 
N_data:88
[click on plot for closeup]
Plot 7
data source:isaac_dark
(QC1 database)
parameter | dataset:qc_darkmed
mean:52.05605ADU
rel. thresholds: 3σ51.42869...52.68341ADU
N_data:88
[click on plot for closeup]
Plot 8
data source:isaac_dark
(QC1 database)
parameter | dataset:qc_dark_stdev
mean:0.03105 
fixed thresholds:0.0...0.06 
N_data:88
[click on plot for closeup]
This plot

This is the trending plot for the ISAAC Aladdin QC parameters derived from DARK frames for read mode DoubleCorrHighBias
and two DIT DIT=0.28sec and DIT=0.56sec

These plots describe:
1. RON12 for DIT=0.28sec,
2. RON23 for DIT=0.28sec,
3. Dark current for DIT=0.28sec,
4. RMS of the three raw dark medians for DIT=0.28sec
5. as Fig.1 but for DIT=0.56sec
6. as Fig.2 but for DIT=0.56sec
7. as Fig.3 but for DIT=0.56sec
8. as Fig.4 but for DIT=0.56sec

RON is derived twice, once from the the first pair of raw frames (RON12 = black points), and once from the second pair of raw frames (RON23 = green points). DoubleCorr detector read modes (= ResetReadRead) show no significant difference between RON12 and RON23.
The dark is the median of the MASTER_DARK product.
The RMS is the standard deviation derived from the three raw frame median values and is a measure of the short-term dark stability for the given DIT.
Find more information about ISAAC DARK frames here.

Other QC parameters related to detector properties (like structural noise, linearity) are trended.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up