[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+R to enforce refresh of scores and dates]
HAWKI trending system: HISTORY plot
Last update: 2011-01-15T07:09:32 (UT)
same group: LINEARITY
close window history: 2009   2010   2011   2012 2013 2014 2015 2016 2017 2018 2019 2020   HEALTH FULL
  ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
template not yet clean, results based on corrected fits frame headers.
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
4 QC1DB MEDIAN 0.73 ADU OFF± 0.4 7 qc_conad this | last_yr | all detector gain, chip 1
5 QC1DB MEDIAN 0.47 ADU OFF± 0.4 7 qc_conad this | last_yr | all detector gain, chip 2
2 QC1DB MEDIAN 0.66 ADU OFF± 0.4 7 qc_conad this | last_yr | all detector gain, chip 3
1 QC1DB MEDIAN 0.41 ADU OFF± 0.4 7 qc_conad this | last_yr | all detector gain, chip 4
3 QC1DB MEDIAN 0.5475 ADU OFF± 0.4 8 qc_conad this | last_yr | all Means of the gain of the four chips
6 QC1DB MEDIAN 0.13549 ADU VAL0.0,0.3 8 qc_conad this | last_yr | all Standard deviations of the measured gain of the four chips
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 4
data source:hawki_detlin
(QC1 database)
parameter | dataset:qc_conad
median:0.73ADU
rel. thresholds: 0.33...1.13ADU
N_data:7
[click on plot for closeup]
Plot 5
data source:hawki_detlin
(QC1 database)
parameter | dataset:qc_conad
median:0.47ADU
rel. thresholds: 0.07...0.87ADU
N_data:7
[click on plot for closeup]
Plot 2
data source:hawki_detlin
(QC1 database)
parameter | dataset:qc_conad
median:0.66ADU
rel. thresholds: 0.26...1.06ADU
N_data:7
[click on plot for closeup]
Plot 1
data source:hawki_detlin
(QC1 database)
parameter | dataset:qc_conad
median:0.41ADU
rel. thresholds: 0.01...0.81ADU
N_data:7
[click on plot for closeup]
Plot 3
data source:hawki_detlin
(QC1 database)
parameter | dataset:qc_conad
median:0.5475ADU
rel. thresholds: 0.1475...0.9475ADU
N_data:8
[click on plot for closeup]
Plot 6
data source:hawki_detlin
(QC1 database)
parameter | dataset:qc_conad
median:0.13549ADU
fixed thresholds:0.0...0.3ADU
N_data:8
[click on plot for closeup]
This plot

Trending report for detector gain in ADU per electrons (ADU/e-).

QC parameter is derived from each detector of the monthly acquired detector mmoitor template.

Keep in mind, that purpose of this plot is montoring only. The actual gain values are unreliable.

Name:
Fits header: QC.GAIN,
QC1 DB: qc_gain

Description:
Photon transfer curve analysis.

Setup:
INS.FILT1.NAME=Ks

These plots describe:
Fig. 1. chip4: The QC parameter for chip #4.
Fig. 2. chip3: The QC parameter for chip #3.
Fig. 3. Average: The QC parameter averaged over all detectors. This view is sensitive to common changes.
Fig. 4. chip1: The QC parameter for chip #1.
Fig. 5. chip2: The QC parameter for chip #2.
Fig. 6. Scatter View: The standard deviation of the QC parameter, obtained from the sample of all detectors. This view is sensitive changes of individual detectors.

Note that the used definition of units are (following DICD):

[GAIN] = ADU/e-
[CONAD] = e-/ADU
 


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up