[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
CRIRES trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2014-07-20T03:47:12 (UT)
same group: flat Hal_IR   ThAr
close window history: 2008 2009 2010 2011 2012 2013   2014   HEALTH
  ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 2.54e+03 ADU/sec VAL800,5000 19 mean_flux this | last_yr | all Halogen lamp, chip 1, WREF=2199.1
2 QC1DB MEDIAN 2.52e+03 ADU/sec VAL800,4500 19 mean_flux this | last_yr | all Halogen lamp, chip 2, WREF=2199.1
3 QC1DB MEDIAN 1.73e+03 ADU/sec VAL500,3500 19 mean_flux this | last_yr | all Halogen lamp, chip 3, WREF=2199.1
4 QC1DB MEDIAN 1.26e+03 ADU/sec VAL300,2500 19 mean_flux this | last_yr | all Halogen lamp, chip 4, WREF=2199.1
5 QC1DB MEDIAN 137 ADU/sec none  19 mean_flux this | last_yr | all IR emitter, chip 1, WREF=2199.1
6 QC1DB MEDIAN 137 ADU/sec none  19 mean_flux this | last_yr | all IR emitter, chip 2, WREF=2199.1
7 QC1DB MEDIAN 97.7 ADU/sec none  19 mean_flux this | last_yr | all IR emitter, chip 3, WREF=2199.1
8 QC1DB MEDIAN 74.2 ADU/sec none  19 mean_flux this | last_yr | all IR emitter, chip 4, WREF=2199.1
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:2.54e+03ADU/sec
fixed thresholds:800...5000ADU/sec
N_data plotted:19
[click on plot for closeup]
Plot 2
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:2.52e+03ADU/sec
fixed thresholds:800...4500ADU/sec
N_data plotted:19
[click on plot for closeup]
Plot 3
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:1.73e+03ADU/sec
fixed thresholds:500...3500ADU/sec
N_data plotted:19
[click on plot for closeup]
Plot 4
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:1.26e+03ADU/sec
fixed thresholds:300...2500ADU/sec
N_data plotted:19
[click on plot for closeup]
Plot 5
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:137ADU/sec
thresholds:none 
N_data plotted:19
[click on plot for closeup]
Plot 6
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:137ADU/sec
thresholds:none 
N_data plotted:19
[click on plot for closeup]
Plot 7
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:97.7ADU/sec
thresholds:none 
N_data plotted:19
[click on plot for closeup]
Plot 8
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:74.2ADU/sec
thresholds:none 
N_data plotted:19
[click on plot for closeup]
This plot

Trending plots monitoring the flux in flat-fields taken with the Halogen lamp or with the IR emitter.

These flats are measured daily with a reference wavelength of 2199.1 nm. The exposures are taken for QC purposes; only the Halogen lamp is also used in other flat-field measurements that are intended for science calibration.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up