method value unit method value
1 QC1DB MEDIAN 3595.88452 ADU/sec VAL800,4500 79 mean_flux this | last_yr | all Halogen lamp, chip 1, WREF=2199.1
2 QC1DB MEDIAN 3354.74683 ADU/sec VAL800,4500 79 mean_flux this | last_yr | all Halogen lamp, chip 2, WREF=2199.1
3 QC1DB MEDIAN 2339.19116 ADU/sec VAL500,3000 79 mean_flux this | last_yr | all Halogen lamp, chip 3, WREF=2199.1
4 QC1DB MEDIAN 1267.76978 ADU/sec VAL300,2500 79 mean_flux this | last_yr | all Halogen lamp, chip 4, WREF=2199.1
5 QC1DB MEDIAN 194.67264 ADU/sec none   79 mean_flux this | last_yr | all IR emitter, chip 1, WREF=2199.1
6 QC1DB MEDIAN 185.34085 ADU/sec none   79 mean_flux this | last_yr | all IR emitter, chip 2, WREF=2199.1
7 QC1DB MEDIAN 133.84702 ADU/sec none   79 mean_flux this | last_yr | all IR emitter, chip 3, WREF=2199.1
8 QC1DB MEDIAN 74.73213 ADU/sec none   79 mean_flux this | last_yr | all IR emitter, chip 4, WREF=2199.1
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:crires_flat
(QC1 database)
parameter | dataset:mean_flux
median:3595.88452ADU/sec
fixed thresholds:800.0...4500.0ADU/sec
N_data:79
[click on plot for closeup]
Plot 2
data source:crires_flat
(QC1 database)
parameter | dataset:mean_flux
median:3354.74683ADU/sec
fixed thresholds:800.0...4500.0ADU/sec
N_data:79
[click on plot for closeup]
Plot 3
data source:crires_flat
(QC1 database)
parameter | dataset:mean_flux
median:2339.19116ADU/sec
fixed thresholds:500.0...3000.0ADU/sec
N_data:79
[click on plot for closeup]
Plot 4
data source:crires_flat
(QC1 database)
parameter | dataset:mean_flux
median:1267.76978ADU/sec
fixed thresholds:300.0...2500.0ADU/sec
N_data:79
[click on plot for closeup]
Plot 5
data source:crires_flat
(QC1 database)
parameter | dataset:mean_flux
median:194.67264ADU/sec
thresholds:none 
N_data:79
[click on plot for closeup]
Plot 6
data source:crires_flat
(QC1 database)
parameter | dataset:mean_flux
median:185.34085ADU/sec
thresholds:none 
N_data:79
[click on plot for closeup]
Plot 7
data source:crires_flat
(QC1 database)
parameter | dataset:mean_flux
median:133.84702ADU/sec
thresholds:none 
N_data:79
[click on plot for closeup]
Plot 8
data source:crires_flat
(QC1 database)
parameter | dataset:mean_flux
median:74.73213ADU/sec
thresholds:none 
N_data:79
[click on plot for closeup]
This plot

Trending plots monitoring the flux in flat-fields taken with the Halogen lamp or with the IR emitter.

These flats are measured daily with a reference wavelength of 2199.1 nm. The exposures are taken for QC purposes; only the Halogen lamp is also used in other flat-field measurements that are intended for science calibration.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up