UVES ECH_SLIC_BLUE CALSELECTOR calibration cascade

DFOS_OPS | CALCHECKER | CALSELECTOR (current) || select other instrument: OCA
OCA rule set:UVES_raw2master.2007-04-01.RLS | other versions
Start-date:2007-04-01
Type of rule set:CALSELECTOR (raw2master type of association; versioning supported)
Sub-cascades: ALL       ECH_BLUE | ECH_SLIC_BLUE | ECH_RED | ECH_ABS_RED | ECH_SLIC_RED | ECH_SLICABS_RED | MOS_RED
General comment:Echelle mode with image slicers, blue arm; identical to "ECH_BLUE" except for STD and SCI (formally there are also data with absorption cell but they are identical to "ECH_SLIC_BLUE" data)

Mouse over ...
- a RAW_TYPE to see the classification rule (M marks a multiple definition)
- a grouping rule to see the complete grouping rule
- a matching rule (e.g. '1,1') to see the product matching rule.
Product tags etc. might be cut off, in that case mouse over the cell to see the full entry.
EXPOSE: marks those non-SCIENCE RAW_TYPEs which can be exceptionally exposed to CALSELECTOR.

Time match rules:
GT: no validity for this static mcalib defined
T: no validity for this mcalib defined
< PREVIOUS time match
||-7.0/+7.0  calib_plan validity (best mjd_obs time difference) and extended validity defined
(-2.0/+2.0)  calib_plan validity not defined, only extended validity (maximum mjd_obs time difference)
[-2.0/+2.0]  calib_plan validity defined, but no extended validity

bottom | right left     
RAW_TYPEBIAS_BLUE FMTCHK_ECH_BLUE ORDERDEF_ECH_BLUE EFLAT_ECH_BLUE DFLAT_ECH_BLUE WAVE_ECH_BLUE STD_ECH_BLUE SCI_SLIC_ECH_BLUE ACQ_ECH
DO_CLASSBIAS_BLUE ARC_LAMP_FORM_BLUE ORDER_FLAT_BLUE FLAT_BLUE DFLAT_BLUE ARC_LAMP_BLUE STANDARD_BLUE SCI_SLICER_BLUE ACQ_ECH
CATGCALIB CALIB CALIB CALIB CALIB CALIB CALIB   EXPOSE SCIENCE ACQUISITION
Grouping rule TPL_A single single TPL_A TPL_A single single single single
Recipe uves_cal_mbias uves_cal_predict uves_cal_orderpos uves_cal_mflat uves_cal_mflat uves_cal_wavecal uves_cal_response uves_obs_scired none
Index123456789

Calibration cascade:
static mcalibs:master calibs:
1 LINE_REFER_TABLE   1,1<       1,1<      
2 MASTER_FORM_BLUE   0,1<              
6 EXTCOEFF_TABLE             1,1< 1,1<  
7 MASTER_RESPONSE_BLUE               0,1<  
10 FLUX_STD_TABLE             1,1<    
11 RESP_FIT_POINTS_CATALOG             1,1<    
15 LINE_INTMON_TABLE           0,1<      
  16 MASTER_BIAS_BLUE     0,1 ||-7.0/+7.0 0,1 ||-7.0/+7.0   1,1 ||-7.0/+7.0 1,1 ||-7.0/+7.0  
    17 LINE_GUESS_TAB_BLUE       1,1      
    18 ORDER_GUESS_TAB_BLUE 1,1 ||-3.0/+3.0            
      19 ORDER_TABLE_BLUE 1,1 ||-3.0/+3.0 1,1 1,1 1,1 1,1  
        20 MASTER_FLAT_BLUE     1,1 1,1  
          21 MASTER_DFLAT_BLUE     0,1 ||-7.0/+7.0 MASSOC  
            22 LINE_TABLE_BLUE 1,1 1,1  
              23 INSTR_RESPONSE_FINE_BLUE 0,1  
                24 RED_SCI_SLICER_BLUE  

Comments:
                three SLICERs; only one STD asociated since otherwise plenty of BIAS, FLATs etc.  

Analysis (technical consistency checks):
raw_type* (CALIB only) OK OK OK OK OK OK OK n/a n/a
mcalibs** n/a OK OK OK OK OK OK OK n/a
powered by QC [createCalibMap v2.1.1] last update: 2020-06-10T10:41:12
*raw_type: is raw_type used in cascade? (CALIB only)
**mcalibs: are all required mcalibs, per raw_type, somewhere listed as products?
top